Products
A company specializing in the development, production and sales of in vitro diagnostic products.
Product classification
Contact Us
Phone/WeChat:
E-mail:
info@immuno-test.com
Address:
Building 1, 578 No.20 Street, Qiantang District, Hangzhou City, Zhejiang Province
Tuberculosis IgG/IgM Rapid Test
Phone:0571-85368996 | Email:info@immuno-test.com
Abbr.:
Specification:
Specimen:
Assay Time:
Model No.:
HTB068T
Model No.: HTB068T
Assay Time: 10-15 minutes
Specimen: Whole Blood,serum or plasma
Tuberculosis IgG/IgM Rapid Test
Graphic Details
INTENDED USE
The Tuberculosis IgG/IgM Rapid Test is a rapid chromatographic immunoassay for the qualitative detection of IgG and IgM antibodies to Tuberculosis virus in human whole blood, serum, or plasma as an aid in the diagnosis of primary and secondary Tuberculosis infections.
[Product Name] | Tuberculosis IgG/IgM Rapid Test |
[Product No.] | HTB068T |
[Assay Time] | 10-15 minutes |
[Specimen] |
Whole blood, serum, or plasma |
[Principle] | Rapid chromatographic immunoassay |
[Storage Temperature] | 2-30℃, DO NOT FREEZE |
[Usage] | For the detection of Tuberculosis antibodies in whole blood, serum, or plasma. |
[Packing] | 20 tests/box |
[Materials Provided] | Individually packed test devices |
Disposable pipettes |
|
Buffer | |
Package insert |
PRINCIPLE
The Tuberculosis IgG/IgM Rapid Test is a qualitative membrane - based immunoassay for the detection of Tuberculosis antibodies in whole blood, serum, or plasma. This test consists of two components, an IgG component and an IgM component.
TEST PROCEDURE
Allow the test device, specimen, buffer, and/or controls to reach room temperature (15 - 30°C) prior to testing.
1.Remove the test device from the sealed pouch and use it as soon as possible.
2. Place the test device on a clean and level surface.
For Serum or Plasma Specimens:
- Hold the dropper vertically, draw the specimen up to the Fill Line (approximately 5μL), and transfer the specimen to the specimen well (S) of the test device, then add 3 drops of buffer (approximately 90μL) and start the timer.
For Whole Blood (Venipuncture/Fingerstick) Specimens:
- To use a dropper: Hold the dropper vertically, draw the specimen 0.5 - 1 cm above the Fill Line, and transfer 1 drop of whole blood (approximately 10μL) to the specimen well (S) of the test device, then add 3 drops of buffer (approximately 90μL) and start the timer.
- To use a micropipette: Pipette and dispense 10 μL of whole blood to the specimen well (S) of the test device, then add 3 drops of buffer (approximately 90μL) and start the timer.
3. Wait for the colored line(s) to appear. Read results at 10 minutes. Do not interpret the result after 20 minutes.
INTERPRETATION OF RESULTS
IgG POSITIVE: The colored line in the control line region (C) appears, and a colored line appears in test line region G. The result is positive for Tuberculosis virus - specific IgG and is probably indicative of secondary Tuberculosis infection.
IgM POSITIVE: The colored line in the control line region (C) appears, and a colored line appears in test line region M. The result is positive for Tuberculosis virus - specific IgM antibodies and is indicative of primary
IgG AND IgM POSITIVE: The colored line in the control line region (C) appears, and two colored lines should appear in test line regions G and M. The color intensities of the lines do not have to match. The result is positive for IgG & IgM antibodies and is indicative of secondary Tuberculosis infection.
NEGATIVE: Only one colored band appears in the control region (C). No line appears in test line regions G or M.
INVALID: No Control line(C) appears.